Introduction
These days, the increase in global mixing from one country to another country, the spread of infectious diseases become faster and thus it became very serious issue [1]. In order to cope with these issues of global mixing and spread of infectious diseases, the fast and accurate detection of the target disease are important for improving the curative efficiency and preventing the spread of disease. Traditional methods, such as immunological or biochemical identification, are time-consuming and laborious, which are not applicable on-site detection of diseases. In addition, antibody detection methods have limitations; for example, they cannot discriminate previously infected patients from currently infected patients. Hence, the use of DNA-based molecular diagnostic methods in various areas including human health, animal health, and food safety has become more popular [2-5]. Molecular diagnostic approaches including rapid detection of pathogens using polymerase chain reaction (PCR) draw the most attention because of the speed and higher accuracy than the traditional serology-based microbial serotyping [6-9].
Recently, studies on on-site usable technology development are reported [10-12]. In order to be used for on-site detection of diseases, where resources are limited, it will be good if the equipment used are regular electricity-independent and the assay can be completed in a short time. And yet, the onsite detection assay should retain the same specificity and sensitivity which are obtained in the laboratory. In this study, we used battery-operated convection PCR (cPCR) device and nucleic acid lateral flow (NALF) immunoassay for detection of a model pathogen, Salmonella species. cPCR is recently developed method which can perform DNA amplification in ultra-fast speed and has been used for on-site detection of pathogens and meat identification [4,5]. The NALF assay has been recently developed for the detection of foodborne pathogens, and pathogens and gene mutations that underlie human diseases [8,13-16]. We optimized the assay conditions for onsite cPCR-based molecular detection of pathogens in combination of NALF immunoassay which is critical for the on-site usable assay development.
Materials and Methods
1. Bacteria strains and preparation of genomic DNA
The bacterial strains used in the current study were Salmonella Enteritidis, Salmonella Typhimurium (NCCP 14771 and NCCP 12219) from the National Culture Collection for Pathogens (NCCP) at the Korea Center for Disease Control & Prevention (Cheongju, Korea). Bacteria were grown overnight in LB broth (Duchefa Biochemic, Haarlem, The Netherlands) at 37°C with shaking.
Genomic DNA (gDNA) was extracted using DNeasy Blood and Tissue kit (Qiagen, Hilden, Germany) according to the manufacturer’s instructions. DNA concentration was determined using a NanoDrop spectrophotometer (Thermo Scientific, Waltham, MA, USA). gDNA copy numbers were calculated for 1 ng of DNA, based on the molecular weight of doublestranded DNA and chromosomal DNA size (http://scienceprimer. com/copy-number-calculator-for-realtime-pcr), as 1.9 × 105 copies/ng of chromosomal DNA for Salmonella.
2. Convection polymerase chain reaction
The primers used were; for detection of Salmonella spp., Biotin-CACGTCGGGCAATTCGT and FAM-GCTTTCCCTTTCCAGTACGC (SPP primers); for detection of S. Enteritidis, Biotin-GTCAGTGCCATACTTTTAATGACTGC and FAM-GTACTATGTCGATACGGTGGGT (SE primers); for detection of S. Typhimurium, Biotin-GCTGTATTTGTTCACTTTTTACCCCT and Digoxigenin-ACCCTGACAGCCGTTAGATATTC (ST primers).
The cPCR mixture (20 μL) contained 1 × PalmTaq HS buffer (Ahram Biosystems, Inc., Seoul, Korea) (supplemented with 1.5 mM MgCl2), 0.2 mM deoxyribonucleotide triphosphates, 0.4 U PalmTaq high-speed DNA polymerase (Ahram Biosystems), and primers for single or multiple PCR detection. For Salmonella detection using single-line NALF (NAD1), SPP primers were used (singleplex detection). To differentiate between S. Enteritidis and S. Typhimurium using double-line NALF (NAD2), ST primers and/or SE primers were used (duplex detection). The concentration of primers used for each experiment was 10 μM each, otherwise stated in the text. Generally, 1.6 ng (3 × 105 copies) of Salmonella gDNA was used as a template. PCR was performed using a battery-operated convection thermal cycler Palm PCR device (G2-12; Ahram Biosystems). The speed was set to T1 (30 cycles in 21 minutes), and the annealing temperature was set to 56°C. All experiments were performed at least in triplicate.
3. Detection of polymerase chain reaction amplification products by electrophoresis and nucleic acid lateral flow analysis
Upon completion of the cPCR, an aliquot of the PCR mixture was analyzed by 1.5% agarose gel electrophoresis (25 minutes at 135 V). PCR products were visualized with ethidium bromide staining and an imaging system (Ultra-Lum, Carson, CA, USA). The NAD1 and NAD2 cassettes (Ahram Biosystems) were designed to indirectly detect amplified PCR products using antibodies against primer tags, such as FAM, DIG, and biotin. When the NAD cassettes were used to detect DNA amplification, 2 μL of PCR product was diluted with 100 μL of the NAD buffer. The diluted solution was added to the reservoir of the NAD cassette, and waited for until the lateral flow was completed. Typically, the reaction development took ca. 5 minutes. The assay outcomes were documented by photographing.
Results and Discussion
1. Optimization of convection polymerase chain reaction conditions for single line nucleic acid detection
The purpose of this study is to optimize the experimental conditions of ultra-fast cPCR and NALF detection using NAD cassette which can be used for on-site detection of pathogens for various diseases. Since the color development of NAD cassette is not dependent of the specific amplification of the target DNA but the presence of the conjugation of the antibody with its antigen, there is possibility of having non-specific color development with non-specifically amplified DNA. Many of these non-specific backgrounds come from the redundant primers in the PCR reaction. Therefore, we tried to optimized primer concentrations of Salmonella spp., S. Enteritidis and S. Thyphimurium, in the cPCR (Fig. 1A and 1B). cPCR was performed at speed level T1 for 21 minutes and the annealing temperature was 56℃. When 10 μM of the primers were used with 1.6 ng of gDNA of S. Enteritidis, non-specific band in no template control (NTC) appeared on both agarose gel and NAD1 detection. However, when the primer concentration was reduced to 3 μM (0.3×), no non-specific band was observed. Similarly, the primer concentration of S. Typhimurium was reduced to 3 μM (0.3×) and non-specific DNA band was observed neither in agarose nor NAD1 detection (Fig. 1B).
One of the way to obtain DNA amplification product before the amplified DNA products reach the saturation phase is to reduce the number of PCR cycles or the PCR running time. We reduced the number of PCR cycles from 30 to 10 cycles (equivalent to 21 minutes to 7 minutes of cPCR running time) with an interval of 5 cycles. The other PCR conditions were kept as the same as the experiment described above. As shown in Fig. 2, when cPCR was performed with S. Enteritidis (Fig. 2A) and S. Typhimurium (Fig. 2B) with SPP primers, the intensity of DNA amplification product was gradually reduced as the PCR running time was reduced from 21 minutes to 7 minutes. Non-specific dimeric DNA production was also reduced in the NTC lanes as the cPCR running time was reduced. The same results were obtained with the agarose gel electrophoresis (Fig. 2A and 2B, upper panels) and with NAD cassettes (Fig. 2A and 2B, lower panels). The results obtained with 14 minutes of cPCR running gave the clear appearance of the DNA amplification band with Salmonella DNA template and no detectable band in NTC experiments. The 14 minutes cPCR running was corresponding to the 20 cycles of cPCR.
2. Optimization of conditions for double line nucleic acid detection
Next, we tried to detect both of S. Enteritidis and S. Typhimurium with a NAD2 cassettes. For balanced detection of S. Enteritidis and S. Typhimurium, the ratio of the primers was optimized. As shown in Fig. 3, the concentration of SE primer which detects S. Enteritidis was fixed to 4 μM, and the concentration of ST primer was varied from 10 to 4 μM. cPCR was run for 21 minutes (equivalent to 30 cycles).
The PCR amplification of targets of S. Enteritidis and S. Typhimurim were well balanced on agarose gel analysis where the ratio of two primer sets were 1 : 1 (0.4 μM : 0.4 μM) on agarose gel analysis. However, due to the generation of nonspecific dimer bands, the red lines were appeared on NAD2 cassettes as false-positive. Thus, we reduced the concentration of primer sets while maintaining the ratio of SE and ST primers as 1 : 1. When the primer concentration was 0.2 μM the two DNA amplicons were appeared with almost equal intensity and at the same time no detectable non-specific band was not observed on the agarose gel (Fig. 3B, upper panel). The similar pattern was observed with NAD2 cassettes: two red lines were appeared with Salmonella DNA templates and no band with NTC when both primer set were used at concentration of 0.2 μM (Fig. 3B, lower panel).
Next, we tried to reduce the PCR running time for the duplex cPCR. As shown in Fig. 4, the cPCR running time was reduced from 21 minutes to 7 minutes (corresponding to 30 cycles to 10 cycles of cPCR). The concentration of the primer was adjusted to have balanced red-line appearance with NAD2 cassettes. The concentration gave the best results were 8 μM of SE primers and 10 μM of ST primers (data not shown). Duplex cPCR amplification was successfully completed with various cPCR running time. When cPCR was performed for 14 minutes (20 cycles), amplified duplex DNA bands appeared on the agarose gel analysis and two red lines were appeared on T section of NAD2 cassette (Fig. 4). At the same conditions, NTC control showed no DNA amplification on the agarose gel analysis and no red line in NAD2 cassette.
Our data suggested that the combination of cPCR [4,5] and lateral flow immunoassay [8,13-16] using NAD cassettes can be applied for a pathogen detection in field. The cPCR conditions to be used with NAD cassettes were found to be optimized by reduction of the primer concentration to 3 μM for singleplex or 2 μM for duplex amplification, or reduction of cPCR running time to 14 minutes which is equivalent to 20 cycles. The assay developed here can be applied to the detection of oral pathogens including Aggregatibacter actinomycetemcomitans, Porphyromonas gingivalis , Prevotella intermedia, Treponema denticola, and Fusobacterium nucleatum both in clinics and laboratory.